 | Press the lift button and the sample will slowly lower into
the magnet. |
 | Press spin, this will spin the sample. You will know the
sample is up to speed when the spin numbers read 20 or 21. |
 | Check to see if the lock signal (peak) is
centered on the screen. If not you'll have to adjust the field
(click field and use the black knob). If you are
unsure of how to do this find Dr. Timm. |
 | Auto-lock: Press auto-lock in order to lock the signal. You
will know the signal is locked when the console numbers go back to the original spin
numbers and the orange line appears at the top of the page. |
 | Once the signal is auto-locked you must lock the signal by
pressing the lock button. |
 | The signal of the sample must be shimmed in order to get
the best signal possible.
Shimming Procedure:
|
 | Read in a Shim file by typing RSH_LAST2004.990
followed by RETURN. |
 | Press the lock-gain button, this gives you the ability to
move the orange raster (line) down |
 | Lower the orange raster by spinning the black knob --- move
it down approximately two blue grid lines |
 | Now press Z, then use the black knob to
move the orange line up and down on the screen: the goal is to get the line as high as
possible on the screen. |
 | Once it is as high as possible hit the Z2
button and repeat the procedure. You should go back and forth between
Z and Z2
as many times as is necessary to get it right but it should only require a few. |
 | Standby: There is a button labeled standby that you need to
press in order to get out of Z (shim) mode. |
 | Then hit the lock-gain button again and spin the
black knob
until the orange line is at the very top of the screen.
|